The innoSens 750 ozone electrode is based on a coated two-electrode system and is suitable for normal water conditions without surfactants. Applications: tap water, drinking water, mineral water, service water, process water.
Features
specifications
Measuring Principle: Membrane covered amperometric
Measuring parameter:Ozone
Measuring range:0.005-2.000/0.05-20.00mg/L(ppm)
Repeatability:± 5%
Slope drift:about <-1%/month (not calibrated)
Working temperature:0-45°C
Compensation:built-in temperature sensor
Working pressure:Max 0.5bar
Flow rate:200-500ml/min
pH range:2-11pH ,4-9pH is recommended for pure water.
Polarisation time:approx. 1 hour for initial use
Zero adjustment:not required
Slope calibration:single point calibration according to DPD method
Material:UPVC, ABS
Dimensions:Outer diameter 25mm, length 220mm
Storage:electrode/membrane cap: 5-40°C,electrolyte:5-35°C
Interference:Residual chlorine (almost negligible), chlorine dioxide
Maintenance interval:membrane cap replacement cycle 1 year, electrolyte replacement cycle 3-6 months
Order No. Description
35-0750-01 innoSens 750-R1,0.005-2.000ppm
35-0750-02 innoSens 750-R2,0.05-20.00ppm
50-0700-30 PA-730 flow cell,acrylic material