The Flumsys 10SC streaming current can be equipped with a pretreatment system to ensure the long-term trouble-free operation of the instrument, with continuous measurement, automatic cleaning,PID control function can be connected to the existing dosing system and start automatic dosing control. The amount of flocculant will be automatically adjusted.
Measurement parameters
Streaming Current Detector (SCD)
Applications
Waterworks、Sewage treatment plant、Sludge dewatering、Need to add flocculant process
Water purification、Sludge dewatering、
Wastewater treatment、Processes requiring
Advantage
Automatic control of flocculant dosing
Flocculant cost savings
Guaranteed water quality
Low operation and maintenance costs
Flumsys 10SC provides water treatment plant operators with an effective tool to optimise and control the amount of flocculant and polymer used!
Water Quality Requirements
Conductivity:< 3000μS/cm
pH: 4~11pH (pH < 7 after controlled flflocculant application, SCD measurement is best)
TSS: < 1000mg/L
● Simultaneous display of actual SC values and relative SC values
● Real-time SC trend graph
● Automatic cleaning function (optional cleaning solenoid valve)
● PID control function
● SC 4-20mA and PID 4-20mA outputs
● 2 high/low alarm outputs
● RS485 Modbus RTU communication
● Password protection against unauthorised operation
● Data logging function, supports U disk to export (Excel)
● Two modes of automatic/manual control
● Optional pre-treatment system for extended maintenance intervals
Technical parameters
Measuring parameter:Streaming Current/SCD,pH
Measurement parameters:Streaming Current
Measurement range:-1000~1000SC
Accuracy:±0.1%
Repeatability:±0.1%
Response time:1s
Operating temperature:Operating temperature:0-50°C
Power supply:220VAC,50/60Hz
Display:7"Touch screen LCD display
Analogue output:SC 4-20mA and PID 4-20mA output, Max. 500Ω
Communication: RS485 Modbus RTU
Alarm relay:High/low alarm contact output, 24VDC/1A
Automatic cleaning:cleaning interval: 0-9999min, cleaning time: 0-999s
Data storage:Real-time data recording, support U disk export (Excel format)
Sampling requirements:flocculant dosing point to sensor time about 3 ~ 5min
Flow rate requirements:1 ~ 4L/min
Protection class: Controller: IP65, Sensor: IP54
Dimensions: Controller: 300x350x200mm, Sensor: 250x350x150mm
Weight: Controller: ca. 10Kg, Sensor: ca. 10Kg
Order Guide
Order No. |
Description |
33-5510-10 |
Flumsys 10SC Streaming Current Detector |
33-5510-11 |
Flumsys 10SC cleaning solenoid valve |
33-5510-12 |
Flumsys 10SC matching filter |
33-5510-10 |
Flumsys 10SC PTFE kit |