The PACON 2501 on-line residual chlorine analyser is an accurate, cost-effective and low-maintenance instrument for continuous on-line monitoring of residual chlorine. It detects the concentration of residual chlorine using the DPD colourimetric method and automatically adds reagents for colourimetric measurement, making it suitable for residual chlorine measurement during chlorination and disinfection and for monitoring residual chlorine concentrations in drinking water networks. Selecting total chlorine reagent, total chlorine concentration can be monitored online.
residual chlorine
total chlorine
piped water
Drinking water network
reverse osmosis
chilled water
swimming pools
shampoo (liquid)
Sterilisation process
DPD colourimetric method for more accurate and stable measurements
- National standard methods, consistent with laboratory methods
Auto-diagnostics and auto-calibration
- Automatic continuous self-diagnostic measurements
- Built-in calibration curve
Low reagent consumption and easy replacement
- 2500/5000 measurements per set of reagents
- Simple reagent change, just pour the DPD powder into the indicator solution to dissolve and use.
Automatic and manual measurement modes
- One-touch (down key) quick start for one measurement
- Automatic measurement period selectable: 60-3600s
Analysis cycle time approx. 2.5 minutes
Measurement time (including cleaning time) is 2.5min.
IP65 protection rating
Indoor wall mounting
4-20mA output
4-20mA output as standard with a maximum load of 500Ω.
RS485 Modbus communication
Standard Modbus RTU communication protocol with 9600 baud rate
Password protection against unauthorised operation
Password: 0022, Chinese/English menu
Technical parameters
Measuring principle: DPD colourimetric method
Measuring range: 0.00-5.00 mg/l (ppm) residual chlorine
Resolution: 0.01 mg/l (ppm)
Accuracy: 1 per cent f.s.
Cycle time: adjustable: 60 - 3600s (system default 300s)
Display: 4.3" LCD touch screen display (measured value, measurement mode and relay status)
Language: Chinese/English
Power supply: 90-260VAC,50/60Hz
Analogue output: 4-20mA output,Max.500Ω
Digital output: RS485 Modbus
Alarm: 2 sets of independently set Hi/Lo alarm points with hysteresis setting, 5A/250VAC
Operating conditions: Operating temperature: 0-50°C; Humidity: 10-95 per cent, non-condensing
Recommendation: Flow rate: 1/min; Pressure: 1 bar
Protection class: IP65
Installation: wall mounting
Size: 300x400x180mm
Order No. |
Descriptions |
33-2501-00 |
PACON 2501 On-line Residual/Total Chlorine Analyser |